Setra Super Count
TM
Operator's Manual 41
Appendix III
Calibration
The Setra Super Count is highly stable, and requires infrequent calibration. While it may be calibrated
as often as desired, the factory calibration is adequate for most counting applications. For precise
weighing applications, or for APW entry via bar code or computer, periodic calibrations may be desir-
able. These calibrations can be done quickly and easily using the scale keyboard and a high accu-
racy Class F or Class S stainless steel weight.
CALIBRATING YOUR SCALE
To calibrate your scale (for a calibration weight contact your dealer), use the following procedure:
1. Remove all objects from the pan.
2. Level the scale.
3. Press the reset key.
4. Press the units key until the scale reads in the units of the known weight.
5. Place the known weight on the pan.
6. Enter the value of the known weight.
7. Press the setup/cal key.
NOTE: To ensure maximum accuracy, a weight should be used which is as close as possible to the
capacity of the scale.